Article number: 238462

SYMPTOMS

  • No communication with TQD

ENVIRONMENT

  • ACQUITY TQD system
  • MassLynx

CAUSE

The Ethernet switch is OFF.

FIX or WORKAROUND

  1. Close MassLynx.
  2. Turn ON the Ethernet switch.
  3. Reboot the acquisition PC.
  4. Open MassLynx.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.