BioAccord fluidics levels drop when not in use - WKB235688
Article number: 235688
SYMPTOMS
- The system is not in use, yet the solvent level in the bottles decreases
ENVIRONMENT
- BioAccord
- ACQUITY RDa
- waters_connect/UNIFI
CAUSE
Evaporation
FIX or WORKAROUND
Fit a bottle seal that has been designed to prevent this (part number 700013742 for RDa Sample Bottle Seal Kit).
ADDITIONAL INFORMATION
id235688, RDA