BioAccord fluidics levels drop when not in use - WKB235688

Article number: 235688

SYMPTOMS

  • The system is not in use, yet the solvent level in the bottles decreases

ENVIRONMENT

  • BioAccord
  • ACQUITY RDa
  • waters_connect/UNIFI

CAUSE

Evaporation

FIX or WORKAROUND

Fit a bottle seal that has been designed to prevent this (part number 700013742 for RDa Sample Bottle Seal Kit).

 

ADDITIONAL INFORMATION

id235688, RDA