Resolution has degraded on a SYNAPT G2-Si - WKB23567

Article number: 23567

SYMPTOMS

  • Resolution in High Resolution mode has dropped from 40k to 25k in positive and negative ion modes

ENVIRONMENT

  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-XS

CAUSE

Resolution has drifted. Time to tune up.

FIX or WORKAROUND

  1. Start infusing leucine enkephalin on the LockSpray or reference probe.
  2. In IntelliStart, under the MS Configuration Mode, run Resolution Optimization. 
  3. Rerun the Resolution Optimization until there is no further improvement in resolution.
  4. Repeat the previous steps for each resolution mode.

ADDITIONAL INFORMATION