Xevo source temperature does not reach the set point - WKB23176

Article number: 23176

SYMPTOMS

  • source temperature does not reach the set point

ENVIRONMENT

  • Xevo TQ-S micro
  • Xevo TQ-S
  • Xevo TQ-XS
  • SYNAPT G2-S
  • SYNAPT G2-Si

CAUSE

Faulty source heater

FIX or WORKAROUND

Replace the source heater (700005347).

ADDITIONAL INFORMATION