Skip to main content
Waters

How to perform the BioAccord cell culture media system suitability test - WKB231508

Article number: 231508

OBJECTIVE or GOAL

Routinely perform the System Suitability Test (SST) to monitor BioAccord system performance.

Some critical quality parameters are checked while performing the SST

  • Solvent background noise level
  • Stability of retention time
  • Stability of response across injections
  • Mass accuracy of identified compounds

ENVIRONMENT

  • BioAccord
  • ACQUITY Premier
  • waters_connect
  • Cell culture media analysis
  • Bioprocess

PROCEDURE

1) Introduction

The SST is designed to ensure that the complete LC-MS system is fully operational prior to routine use. This test is designed so that stringent quality criteria (mass accuracy, RT stability, peak separation, peak response reproducibility, and solvent background) are assessed using a Waters-certified standard solution (Vion Test Mix) in a 10-minute run. This will help users to assess the status of their LC-MS instrument prior to performing data acquisition. The method enables screening for nine substances using the mass and retention time as quality criteria. The method can also display and identify matched fragments, in order to further demonstrate the system’s screening capabilities.

Waters recommends that you routinely perform this or a similar test to maintain the highest levels of confidence in your data.

XIC Chromatogram.png
Example of extracted ion chromatogram of the nine injected compounds for positive ion mode.

2)    Checklist

  • Check solvents and prime the ACQUITY Premier LC system and wash syringes.
  • Create a new analysis.
  • Go to initial conditions.
  • Inject two blanks and five injections of the Vion Test Mix.
  • Review the report to ensure that the system meets performance criteria.

CCM SST workflow.PNG

3)    Sample preparation

  • Prepare a blank sample by transferring 1 mL of H2O plus 0.1% FA (mobile phase A) into a Waters vial and place it in the Sample Manager in position 1:A,1.
  • Transfer the content of Waters part 186008462 Vion Test Mix ampoule into a Waters vial and place it in the Sample Manager in position 1:A,2.

4)    Detailed procedure

  1. Check the ACQUITY Premier UPLC solvent bottles and ensure that there is sufficient volume of mobile phase.
  2. Ensure that there is a blank on position 1:A,1 and the Vion Test mix on position 1:A,2.
  3. Confirm that there is sufficient lockmass solution (left bottle of RDa fluidics).
  4. Open UNIFI software.
  5. When performing the SST for the first time, import the UNIFI UEP file.
    • Import the "SST CCM.uep" file to UNIFI (contains sample list, analysis method, example data, and report template).
  6. Create a new analysis by selecting "Create analysis by acquiring new data".
  7. Select the "CCM SST sample list".
  8. Select the "CCM SST Analysis Method", and click Next.
  9. Adjust the analysis name and folder as appropriate, and click Next.
  10. Select the system, click Next, and then click Finish.
  11. Click on the bottom-right green check mark icon to open the "Instrument System" side panel.
  12. On the Binary Solvent Manager, select Prime A/B solvents, select start priming, and close the window.
  13. On the Sample Manager FTN, select Prime Syringe, perform prime, and close the window when complete.
  14. On Summary, select Go to initial conditions.
  15. Wait for the delta pressure of the binary solvent manager to be less than 30 psi.
  16. Select sample list lines to be injected.
  17. Click the Start button (be sure to select "Process samples", "Save results", "Generate a report", and "Save report and do not print".
  18. When acquisition is complete, review the report for Pass or Fail. (see example report: "CCM SST Report.pdf")

SST PASS: Complete

The BioAccord LC-MS system is fully operational. The quality criteria for mass accuracy, RT stability, MS response and reproducibility are within expectations across five replicates injections for each positive and negative polarities. The analysis results meets the following criteria: 1). Identified components mass accuracy is within +/- 5 mDa 2). Observed RT %RSD is better than 1.00% 3). MS Response %RSD is better than 10.0% 4). Sulfadimethoxine mean response is greater than 5'000 counts for positive polarity 5). Sulfadimethoxine mean response is greater than 1'000 counts for negative polarity

SST FAIL: Complete with Error

The BioAccord LC-MS system requires user attention before acquiring new data. One or more quality criteria are not meeting expected results across five replicates injections for each positive and negative polarities. The following page of the report will show red flags where passing criteria have not been achieved.

Here are some tips for troubleshooting the system:

1). Mass error flags will require a new RDa detector setup to be performed.
2). Observed RT %RSD and MS Response %RSD flags indicates an issue on the chromatographic system. This requires to check the flow path for leaks, prepare fresh solvents and prime the system.
3). A flag on sulfadimethoxine response indicates a sensitivity issue with the RDa detector. This may require to make a fresh sample or change the aperture disc.
4). The limit status column of the analysis injection list allows to review which specific injection is causing issues

5)    Analysis method

 

Purpose

Manage Components

         
Component name Expected RT (min) Expected neutral mass (Da) Adducts Formula Concentration (µg/mL)
Sulfaguanidine 2.40 214.0524 +H, -H C7H10N4O2S 5.0
Acetaminophen 3.60 151.0633 +H, -H C8H9NO2 10.0
Val-tyr-val 3.80 379.2107 +H, -H C19H29N3O5 2.5
Caffeine* 4.00 194.0804 +H C8H10N4O2 1.5
Leucine enkephalin 4.60 555.2693

+H, -H

C28H37N5O7 2.5
Verapamil* 5.70 454.2832 +H C27H38N2O4 0.2
Sulfadimethoxine 5.80 310.0736 +H, -H C12H14N4O4S 1.0
Reserpine* 5.90 608.2734 +H C33H40N2O9 0.6
Terfenadine* 6.50 471.3137 +H C32H41NO2 0.2
*The compound does not ionize in negative ion mode.
 

Purpose

Default Amounts

         
Add custom level POS NEG      

 

 

Instrument

Column Chemistry

ACQUITY Premier HSS T3 2.1 x 150 mm (186009469)

   
Column temperature 40 °C

Flow rate

250 µL/min

Mobile phase A

H2O / 0.1% FA

Mobile phase B

90% ACN / 10% IPA / 0.1% FA

Wash

40% ACN / 40% H2O / 20% IPA

Purge

90% H2O / 10% ACN

Seal wash

90:10 water/methanol

 

 

Time (min)

Flow Rate (mL/min)

Composition A (%)

Composition B (%)

Curve

0.00

0.250

95.0

5.0

Initial

0.50

0.250

95.0

5.0

6

7.00

0.250

15.0

85.0

6

7.80

0.250

15.0

85.0

6

8.00

0.250

95.0

5.0

6

10.00

0.250

95.0

5.0

6

 

 

ACQUITY RDa Detector

 

Mode

"Full scan" or "Full scan with fragmentation"

Mass range

Small molecules (50 – 800 m/z)

Polarity

Positive or Negative

Scan rate

5 Hz

Cone voltage

Default (30 V) for positive and default (40 V) for negative

Fragmentation cone voltage

Custom 60 V to 140 V

Events Table

0.00 Divert to waste

1.00 Divert to MS

8.00 Divert to waste

Capillary voltage

Default (1.50 kV) for positive and default (0.80 kV) for negative

Desolvation temperature

Default (550 °C)

Intelligent data capture

On

Lockmass correction mode

Standard

 

 

Sample List Setup  
Sample type Blank or QC
Run Time (min) 10
Injection volume (µL) 0.5
Replicates 1
Polarity (Instrument) Positive / Negative

 

 

Peak Processing Settings

3D Peak Detection

         
Low energy intensity threshold 1000 counts        
 

Targeted Screen Settings

Target by Retention Time

         
Enable screen by retention time Enabled        
Identify target peak using rule Largest in the RT window        
Use absolute retention time identification tolerance 0.5 minutes        
 

Targeted Screen Settings

Target by Mass

         
Target mass tolerance 20 ppm        
Generate predicted fragments from structure Enabled        
Fragment match tolerance 10.0 mDa        
Extract mass chromatogram Enabled        
Tolerance Manual, 30 mDa        
Extract mass chromatogram containing all identified ions Enabled        
 
 

Summary Calculations

         
Field Sample type Group by      
Observed RT (min) QC Sample position      
MS Response QC Sample position      
 
 

Limit Checks

         
Node Field name Component Level Error minimum Error maximum
Component Mass error (mDa) All All -5 5
Summary fields Observed RT (min):% RSD (%) All All   1
Summary fields MS Response:% RSD (%) All POS   10
Summary fields MS Response:% RSD (%) All NEG   10
Summary fields MS Response:Mean Sulfadimethoxine POS 5000  
Summary fields MS Response:Mean Sulfadimethoxine NEG 1000  

Sample type is QC for all limit checks field

 

ADDITIONAL INFORMATION

 

id231508, eluent, isopropanol

Not able to find a solution? Click here to request help.