No Xevo QTof beam after a fresh install of MassLynx on a new PC - WKB22645
Article number: 22645
SYMPTOMS
- No beam
- Instrument responds (API gas turns on, baffle turns, can go into Operate)
ENVIRONMENT
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
- Xevo QTof
- Xevo Tof
- Xevo G2 QTof
- Xevo G2 Tof
- Xevo G2-S
- Xevo G2-XS
- MassLynx
CAUSE
The factory settings need to be imported.
FIX or WORKAROUND
- On the tune page, go into system view (setup drop-down menu > system view). The password is access.
- Select the File drop-down menu and select Factory Defaults Import or Load System Settings (from EPC).
- If there is still no beam, increase the detector V in steps of 50 V (System drop-down menu > auto values > Detector V), and select update. If the previous detector V is known, set it to that voltage.