Low sensitivity on a Xevo TQ-S micro - WKB226464

Article number: 226464

SYMPTOMS

  • Low sensitivity reported with LC-MS and ESI acquisitions
  • Setup solution peaks do not meet resolution requirements.

ENVIRONMENT

  • Xevo TQ-S micro

CAUSE

The cone gas nozzle is not fitted correctly.

FIX or WORKAROUND

Refit the cone gas nozzle, making sure that it sits flush with the ion block assembly.

ADDITIONAL INFORMATION