Analyzer vacuum failure on Xevo TQ-S, although all three turbo pumps are running at 100% - WKB224902
Article number: 224902
SYMPTOMS
- Analyzer vacuum failure error message
- All three turbo pumps show 100% on the Diagnostics tab of the Tune page
- Message regarding high vacuum trip is seen when trying to go into Operate
- On the Vacuum menu of the Tune page, the options include Cancel Pump
ENVIRONMENT
- Xevo TQ-S
CAUSE
Communication issue
FIX or WORKAROUND
- Close MassLynx.
- Restart the MassLynx workstation.
- Log in to Windows but do not start MassLynx yet.
- Reset the instrument electronics using the toggle switches on the back of the Xevo TQ-S:
- Toggle the Vacuum Override (top) switch down. (Note: when toggling these switches, you must pull them toward you, then toggle.)
- Toggle the Electronics switch (bottom) down, Wait for the fans to stop.
- Toggle the Electronics switch back up. Wait for the fans to slow down to their normal speed.
- Toggle the Vacuum Override switch up.
- Open MassLynx.
ADDITIONAL INFORMATION
