At what vacuum level does the RDa go into Operate? - WKB225079

Article number: 225079

ENVIRONMENT

  • RDa
  • BioAccord

ANSWER

The RDa must reach 7e-7 mbar vacuum.

ADDITIONAL INFORMATION

At this point, the vacuum level allows the RDa to go into Operate, but vacuum is not yet low enough for the software to run the setup successfully.

Wait for another hour before running the setup.

id225079, RDA