Xevo TQ-S turbo pumps do not turn on when pumping down - WKB224652
Article number: 224652
SYMPTOMS
- When selecting pump down on the tune page, the rotary pumps turn on but the turbo pumps do not
- Readbacks for the turbo speed are 0 and do not change
- Cooling fans on the turbos do not power-on
ENVIRONMENT
- Xevo TQ-S
CAUSE
The turbo controllers are not getting a pump down signal from the electronics. Something has latched into an error state and cycling the power does not reset the error state.
FIX or WORKAROUND
- Close MassLynx and power-down the electronics.
- Switch the system to pump override and the rotary pumps and turbos should turn on—the turbo cooling fans should also be on.
- Power-on the electronics and establish communications with MassLynx.
- When the tune page is opened, turn off the pump override switch and the system should continue to pump down, and the turbo readbacks should now update normally.
ADDITIONAL INFORMATION