Xevo G2-XS produces only electronic noise during acquisition - WKB220776
Article number: 220776
SYMPTOMS
- Chromatogram of sample probe and LockSpray trace show only electronics noise (ADC noise level)
- UV Trace looks normal
- No signal in MassLynx Tune page on Sample and LockSpray position with LeuEnk
ENVIRONMENT
- MassLynx
- Xevo G2-XS QTof
CAUSE
FIX or WORKAROUND
- Close MassLynx.
- Reset the Xevo G2-XS electronics.
See KCS article WKB56994. - Wait five minutes and reopen MassLynx.
- Confirm that the signal is back with LeuEnk on the Tune page.
ADDITIONAL INFORMATION
id220776, MLYNX, MLYNXV41, SUPMM, XEVOG2XSTF