Xevo G2-XS produces only electronic noise during acquisition - WKB220776

Article number: 220776

SYMPTOMS

  • Chromatogram of sample probe and LockSpray trace show only electronics noise (ADC noise level)
  • UV Trace looks normal
  • No signal in MassLynx Tune page on Sample and LockSpray position with LeuEnk

ENVIRONMENT

  • MassLynx
  • Xevo G2-XS QTof

CAUSE

 

FIX or WORKAROUND

  1. Close MassLynx.
  2. Reset the Xevo G2-XS electronics.
    See KCS article WKB56994.
  3. Wait five minutes and reopen MassLynx.
  4. Confirm that the signal is back with LeuEnk on the Tune page.

ADDITIONAL INFORMATION

 