Retention time shift - WKB21197
Article number: 21197
SYMPTOMS
- Chromatographic peaks shift in time; problem is seen only on the Xevo TQ MS with the Adlink EP
-
On fast scans (e.g., dwell times and ISD of 0.005 s) there may be a problem with the timing of acquired scans
-
Retention times of chromatographic peaks may appear to shift by up to -6 seconds for every minute of run time
-
More likely with longer run times and later eluting peaks—for example, in an eight-minute run a peak seen at 7.2 minutes (with slow scan speeds)
may appear to elute at 6.7 minutes
-
The total time acquiring the run may also be increased (e.g., an 8-minute run may take nearly 9 minutes to acquire)
-
Peaks seen in the ACQUITY Console plots maintain their true retention times
-
There may be more scans acquired than the theoretical maximum for an acquisition
ENVIRONMENT
- Xevo TQ MS
- MassLynx
CAUSE
Bug in software: SCN 876, SCN 905, and SCN 950
FIX or WORKAROUND
- Use a duty cycle (interscan delay + dwell time) of more than 0.100 seconds (100 ms). At this scan speed there is negligible effect on retention times.
- Install patch SCN 948 (part 667005327), which was released to fix this issue.
