Weak and or no signal from Lockspray on tune page - WKB21104

Article number: 21104

SYMPTOMS

  • Lockspray shows very weak/no signal
  • The Lockspray baffle is in the correct position
  • The analyte sprayer does generate ions
  • Fluidics is delivering Lockspray solution to the Lockspray inlet flange

ENVIRONMENT

  • Xevo QToF
  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2-Si
  • Vion

CAUSE

The Lockspray capillary is bent or blocked.

FIX or WORKAROUND

  1. Replace the Lockspray capillary. (Spares kit 700004596)
  2. Verify that ions can be seen when infusing Lockspray solution.

ADDITIONAL INFORMATION

This also happens when the LockSpray capillary tip protrudes too far out and the end of the capillary isn't in the peak union

Replace lockspray assembly (700004588) if replacing capillary doesn't resolve.

Probe tip can be adjusted and purging of fluidics is recommended.

Do not run greater than 20uL/min flow rates thru the lockspray capillary when in use to prevent it from coming loose and protuding too far out of the sleeve. 

id21104, MALDIG2SI, SYNG2SIHD, VION, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOQTOF