No ions from Lockspray on Xevo G2-S ToF tune page - WKB21104

Article number: 21104

SYMPTOMS

  • Lockspray shows no ions
  • The Lockspray baffle is in the correct position
  • The analyte sprayer does generate ions
  • Fluidics is delivering Lockspray solution to the Lockspray inlet flange

ENVIRONMENT

  • Xevo QToF
  • Xevo G2 QToF
  • Xevo G2-S ToF
  • Xevo G2-XS ToF
  • SYNAPT G2-Si

CAUSE

The Lockspray capillary is blocked.

FIX or WORKAROUND

  1. Replace the Lockspray capillary.
  2. Verify that ions can be seen when infusing Lockspray solution.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.