SYNAPT G2-Si IMS-Config tab settings are dimmed - WKB209823

Article number: 209823

SYMPTOMS

  • IMS-Config tab settings are dimmed (unavailable)

ENVIRONMENT

  • SYNAPT G2-Si
  • SYNAPT XS
  • MassLynx

CAUSE

The settings are not enabled on the TriWave tab.

FIX or WORKAROUND

  1. Enable Manual Controls for IMS on the TriWave tab.
  2. Enable the necessary sections of the IMS-Config tab.

ADDITIONAL INFORMATION