No signal seen on Xevo TQD - WKB20952
Article number: 20952
SYMPTOMS
- Capillary Voltage readback is 0 V when in Operate mode
- All other readbacks are okay
- Spray from capillary is okay
- Isolation valve is open
ENVIRONMENT
- Xevo TQD
CAUSE
Unknown
FIX or WORKAROUND
Set a new capillary voltage (different than what is in use), and then set it back to the original value while in Operate mode.
ADDITIONAL INFORMATION
The capillary voltage on Xevo TQD sometimes hangs. Changing the request value resends a request to the PSU.
id20952, XEVOTQD, XEVOTQDIVD