Stop flow input activated with SQ Detector 2 - WKB209551
Article number: 209551
SYMPTOMS
- When in Operate, the green LED shows Operate status is off
- Stop flow signal sent when the gases are all on
- No other errors observed
- Stop flow cable appares to be connected securely on green terminal block connectors at both ends (BSM and MS)
- Disconnect Stop Flow cable from back of BMS, flow will start as expected (temporary fix)
- Reconnect Stop Flow cable to the back of BSM, flow immediately stops, Stop Flow Input message appears.
ENVIRONMENT
- SQ Detector 2
- Xevo TQD
CAUSE
Error on EPC
FIX or WORKAROUND
- Close down MassLynx.
- Restart EPC, either via telnet or by pressing the reset button next to the power button on the front of the instrument.
- Wait until the EPC fully reboots (wait five minutes or until you hear the EPC chime).
- Confirm that the Operate LED is lit after you put the instrument into Operate.
ADDITIONAL INFORMATION
id209551, A-SQ, SQD2