Xevo TQD source pressure test leak trip - WKB20200

Article number: 20200

SYMPTOMS

  • Source pressure test fails
  • Error message: "Source pressure exhaust leak" or "Source pressure leak trip"
  • Source pressure readback remains low during the test (for example, 0 mbar or slightly higher)

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-MS
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo G2-Tof
  • Xevo G2-QTof
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2 MS
  • SYNAPT G2 HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • ACQUITY QDa

CAUSE

Loose bottle waste cap or faulty Waste bottle.

FIX or WORKAROUND

1) Tighten the bottle waste cap.

2) If this does not work replace bottle.

 

 

ADDITIONAL INFORMATION

 