Skip to main content
Waters

Source pressure test leak trip on the QDa, SYNAPT G2, SYNAPT G2-S, SYNAPT G2-Si, Synapt HDMS, Xevo G2, Xevo G2-S, Xevo G2-XS, Xevo TQ MS, Xevo TQD, Xevo TQ-S, Xevo TQ-S Micro or Xevo TQ-XS - WKB20200

Article number: 20200

SYMPTOMS

  • Source pressure test fails
  • Error message: "Source pressure exhaust leak" or "Source pressure leak trip"
  • Source pressure readback remains low during the test (for example, 0 mbar or slightly higher)

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-MS
  • Xevo TQ-S
  • Xevo TQ-XS
  • Xevo TQ-S cronos
  • Xevo TQ-S micro
  • Xevo G2-Tof
  • Xevo G2-QTof
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2 MS
  • SYNAPT G2 HDMS
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • ACQUITY QDa

CAUSE

Loose bottle waste cap or faulty Waste bottle.

FIX or WORKAROUND

1) Tighten the bottle waste cap.

2) If this does not work replace bottle.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.