How to calculate the heat dissipation of your MS instrument - WKB20195

Article number: 20195

OBJECTIVE or GOAL

Calculate the heat dissipation for an MS instrument, including backing pumps.

ENVIRONMENT

  • Single Quadrupole instruments
    ACQUITY QDa Detector
    SQ Detector 2
    ACQUITY SQD
    3100 Mass Detector
    ZQ Mass Detector
    Tandem Quadrupole instruments
    Xevo TQ-XS
    Xevo TQ-S micro
    Xevo TQD
    Xevo TQ-S
    Xevo TQ MS
    TQ Detector (ACQUITY TQD)
    Quattro Premier/Quattro Premier XE
    Quattro micro API
    Quattro micro GC
    Quattro Ultima
    Quattro LC
    ToF (Time of Flight) instruments
    SYNAPT G2-Si
    SYNAPT G2-S
    SYNAPT G2
    SYNAPT
    Xevo G2-S
    Xevo G2-XS
    Magnetic Sector MS
    AutoSpec Premier

PROCEDURE

  1. To calculate the heat dissipation of your MS instrument, consult the power requirements section of your instrument's site prep guide and record the maximum power consumption of all necessary components. This will differ depending on pumping configuration. When reporting the heat dissipation we work on a worst case scenario and assume that the total power consumed is converted to heat. Add together the total power consumption of the MS and backing pumps. This gives you the heat dissipation value in Watts. In documents where this value is reported in VA, Volt-Amperes, 1 VA = 1 W.

 

ADDITIONAL INFORMATION

1 Watt = 3.412142 BTU/hr