How to calculate the heat dissipation of your MS instrument - WKB20195
Article number: 20195
OBJECTIVE or GOAL
Calculate the heat dissipation for an MS instrument, including backing pumps.
ENVIRONMENT
- Single Quadrupole instruments
- ACQUITY QDa Detector
SQ Detector 2
ACQUITY SQD
3100 Mass Detector
ZQ Mass Detector - Tandem Quadrupole instruments
- Xevo TQ-XS
Xevo TQ-S micro
Xevo TQD
Xevo TQ-S
Xevo TQ MS
TQ Detector (ACQUITY TQD)
Quattro Premier/Quattro Premier XE
Quattro micro API
Quattro micro GC
Quattro Ultima
Quattro LC - ToF (Time of Flight) instruments
- SYNAPT G2-Si
SYNAPT G2-S
SYNAPT G2
SYNAPT
Xevo G2-S
Xevo G2-XS - Magnetic Sector MS
- AutoSpec Premier
PROCEDURE
- To calculate the heat dissipation of your MS instrument, consult the power requirements section of your instrument's site prep guide and record the maximum power consumption of all necessary components. This will differ depending on pumping configuration. When reporting the heat dissipation we work on a worst case scenario and assume that the total power consumed is converted to heat. Add together the total power consumption of the MS and backing pumps. This gives you the heat dissipation value in Watts. In documents where this value is reported in VA, Volt-Amperes, 1 VA = 1 W.
ADDITIONAL INFORMATION
1 Watt = 3.412142 BTU/hr