IntelliStart Instrument Resolution fails on a Xevo TQ-XS - WKB198139
Article number: 198139
SYMPTOMS
- IntelliStart Instrument Resolution struggles to pass at Unit Mass Resolution and High Mass Resolution
- IntelliStart Instrument Resolution fails at Low Mass Resolution and Quantitative
- The peak splits in two at low resolution:

- Zooming in to the peak top of one of the peaks on the tune page (ex. 455.3 and span of 0) shows the intensity fluctuating by 30-40% from scan to scan
- Resolution results fail with ridiculous values (0.5 Da for one mass and 0.9 Da for others)

ENVIRONMENT
- Xevo TQ-XS
- 186008719 setup solution for the Xevo TQ-XS
- IntelliStart
CAUSE
Unstable spray due to kink or blockage in the capillary
FIX or WORKAROUND
Replace the capillary: