What are the exhaust requirements for a Xevo TQ-XS? - WKB197869
ENVIRONMENT
- Xevo TQ-XS
- Xevo G2-XS
ANSWER
A brief overview includes the following details:
The exhaust line supplied with the system must either be fed into a ducted laboratory fume hood or connected to a laboratory exhaust system.
The laboratory exhaust system must be capable of supporting a gas load of 2000 L/hour. (The pressure within the laboratory exhaust must be less than atmospheric pressure, but not less than -10 mbar gauge, while under gas load.)
The backing pump exhaust gases must be vented to the atmosphere outside the laboratory using a user-supplied fume hood or industrial vent. (The fume hood or industrial vent must be equipped with an extraction fan system to enable adequate displacement of the exhaust gases.)
ADDITIONAL INFORMATION
More details on laboratory exhaust, source exhaust, and backing pump exhaust requirements are in the Site Preparation Guide for the Xevo TQ-XS. Review the Waters Xevo TQ-XS Site Preparation Guide (715005172); the Exhaust Outlets section begins on page 18 in Version 4 of this document.