Pump down fails on the SYNAPT G2-S - WKB19778
Article number: 19778
SYMPTOMS
- Does not pump down
- Turbo timeout
ENVIRONMENT
- SYNAPT G2-S
CAUSE
Air supply to the pump control box is turned off. The supply should be regulated at between 2.8 and 4.2 bar (40 and 60 psi).
FIX or WORKAROUND
Restore the air supply and pump down.
ADDITIONAL INFORMATION
The air is supplied to pneumatic valves in the backing lines. The turbos cannot spin up if the backing pump is not pulling on the outlet.