How do I replace the diaphragm backing pump on a standard ACQUITY QDa Mass Detector? - WKB197736
Article number: 197736
OBJECTIVE or GOAL
Replace a faulty diaphragm backing pump on a standard ACQUITY QDa Mass Detector or RADIAN ASAP Direct Mass Detector
ENVIRONMENT
- ACQUITY QDa Mass Detector
- RADIAN ASAP Direct Mass Detector
PROCEDURE
- Turn off the QDa Mass Detector/RADIAN ASAP Direct Mass Detector.
- Remove the 6-mm o.d. tubing from the 6-mm elbow and the pump power and control connection at the back of the detector, as highlighted in the picture.
- Gently lift the diaphragm pump upwards off the bracket that secures it to the back of the QDa Mass Detector
- Mount the new/refurbished diaphragm pump on the QDa.
- Reinstall the 6-mm o.d. tubing and pump power and control connection at the back of the detector.
- Turn on the QDa.