No capillary voltage readback on Xevo TQD - WKB196110
Article number: 196110
SYMPTOMS
- No capillary voltage readback on Xevo TQD
ENVIRONMENT
- Xevo TQD
- MassLynx
- Xevo TQ Absolute
CAUSE
Unknown
FIX or WORKAROUND
- Set the Xevo TQD to Standby and turn off the API gas.
- Disconnect the sample probe HV cable.
- Reseat the probe in the source enclosure and reconnect the sample probe HV cable.
ADDITIONAL INFORMATION
The power supplies are designed to switch off if they register a certain number of spikes per second Or the capillary is too close to the cone.
This can be reset by going to Operate > Standby > Operate and seeing if the capillary voltage returns.
If it does, it's then a case of monitoring to see if it was a one-off or a problem elsewhere.
id196110, XEVOG2XSQT, XEVOTQ, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD