Low sensitivity on Xevo TQD - WKB196191
Article number: 196191
SYMPTOMS
- Drop in sensitivity after switching to a new assay
- Drop in sensitivity after switching source enclosure
- No issues found in LC delivery
ENVIRONMENT
- Xevo TQD
- Xevo TQ-S micro
CAUSE
The probe must be re-aligned for the new assay.
FIX or WORKAROUND
- Infuse a known standard.
- Adjust the length of the capillary, the tilt, and the angle of the probe to maximize signal.
ADDITIONAL INFORMATION
id196191, XEVOTQD, XEVOTQDIVD, XEVOTQSMIC