Low sensitivity on Xevo TQD - WKB196191

Article number: 196191

SYMPTOMS

  • Drop in sensitivity after switching to a new assay
  • Drop in sensitivity after switching source enclosure
  • No issues found in LC delivery

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ-S micro

CAUSE

The probe must be re-aligned for the new assay.

FIX or WORKAROUND

  1. Infuse a known standard.
  2. Adjust the length of the capillary, the tilt, and the angle of the probe to maximize signal.

ADDITIONAL INFORMATION

 

id196191, XEVOTQD, XEVOTQDIVD, XEVOTQSMIC