The CTCPAL3 HS injections are not in sequence - WKB194113
Article number: 194113
SYMPTOMS
-
The first injection is at vial 1, and the second injection is at vial 11
ENVIRONMENT
-
CTC PAL3
CAUSE
The R60 tray default configuration is in the vertical order, but the row direction is set in the inlet method, which is in the status page.
FIX or WORKAROUND
Change the row direction to the Column direction.
ADDITIONAL INFORMATION
id194113,