The CTCPAL3 HS injections are not in sequence - WKB194113

Article number: 194113

SYMPTOMS

  • The first injection is at vial 1, and the second injection is at vial 11

ENVIRONMENT

  • CTC PAL3

CAUSE

The R60 tray default configuration is in the vertical order, but the row direction is set in the inlet method, which is in the status page.

FIX or WORKAROUND

Change the row direction to the Column direction.

clipboard_e1623af631b121c7c53b3cfa294ee70f5.png

clipboard_ee229a7483d5da5adcfb778fa0754d204.png

ADDITIONAL INFORMATION

 