Low sensitivity in negative mode for lockspray and toxicology screening on Xevo G2-XS - WKB193985
Article number: 193985
SYMPTOMS
- Lockspray beam is a few thousand counts in negative for leucine enkephalin
- Lockspray beam is okay in positive mode for leucine enkephalin
- Samples are okay in positive mode but very poor in negative
ENVIRONMENT
- Xevo G2-XS
CAUSE
Arcing of the reference probe.
The reference probe is or has been partially blocked, and the capillary is protruding more than it should, causing arcing that affects negative mode specifically.
FIX or WORKAROUND
Rebuild the reference probe.
ADDITIONAL INFORMATION
700004674 - ESI REFERENCE PROBE SERVICE KIT
id193985, Electrospray, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF