Randomly occurring "low desolvation gas flow" error on the Xevo TQ-XS - WKB194299

Article number: 194299

SYMPTOMS

  • Randomly occurring "Low desolvation gas flow" error message
  • Nitrogen supplied via liquid dewars
  • Pressure at regulator: 100 psi
  • Source pressure test successful

ENVIRONMENT

  • All Xevo TQ type mass spectrometers
  • All Xevo Tof/QTof type mass spectrometers (G2, G2-S, G2-XS)
  • SQ Detector
  • TQ Detector
  • SQ Detector 2

CAUSE

Nitrogen supply dewar being changed

FIX or WORKAROUND

After changing the nitrogen supply dewar, reset the mass spectrometer.

ADDITIONAL INFORMATION

 

id194299, A-SQ, A-SQ, CYCIMS, MALDIG2SI, MRT, Q3100MSD, SQD, SQD2, SYNAPT, SYNAPTMS, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, TQD, TQDIVD, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSQT, XEVOG2XSTF, XEVOG3QTOF, XEVOQTOF, XEVOTQ, XEVOTQA, XEVOTQA, XEVOTQD, XEVOTQDIVD, XEVOTQIVD, XEVOTQS, XEVOTQSCRO, XEVOTQSIVD, XEVOTQSMIC, XEVOTQXS, XEVTQSMIVD, XEVTQXSIVD