Pump down fails on SQD, TQD or Xevo TQ-S micro - WKB19184

Article number: 19184

SYMPTOMS

  • Turbo Speed reaches only 6%
  • No audible sound comes from rotary pump
  • Outlets that feed the rotary pump have the correct voltages
  • Rotary pump power cable is ok and fuse is intact

ENVIRONMENT

  • Xevo TQ-S micro
  • TQD
  • SQD
  • Rotary Pump
  • Leybold Sogevac SV40BI

CAUSE

Rotary pump failure

FIX or WORKAROUND

Replace the rotary pump.

ADDITIONAL INFORMATION

Sogevac SV40BI part number 186006347