Noise increases for trichloroacetic acid only among the haloacetic acids - WKB191242

Article number: 191242

SYMPTOMS

  • Noise increases for trichloroacetic acid only among the haloacetic acids.
  • It rises before the peak.
  • The longer the capillary protrudes out of the probe tip, the lower the noise.

ENVIRONMENT

  • TQD UPLC

CAUSE

Dirt on the probe tip

FIX or WORKAROUND

  1. Replace or clean the probe tip and replace the capillary tube with a new one.