How to reset Typhoon communications with Vion IMS QTof - WKB1901
Article number: 1901
OBJECTIVE or GOAL
Reset Typhoon communications with Vion when it is unresponsive (i.e., non-storage display is not updating).
ENVIRONMENT
- Vion IMS QTof
- UNIFI 1.8
PROCEDURE
There are two ways to reset Typhoon communications with Vion:
1)
- In UNIFI, go to either the Instrument System or the Analysis tab.
- Bring up the "Instrument System Status" panel (green check mark icon at the bottom-right of the screen).
- Click the Vion IMS QTof drop-down tab.
- Click Reset instrument.
2)
- In UNIFI, go to Instrument System.
- In the Console Navigation panel, select "Vion IMS QTof" in the "System" menu.
- Select "Setup" in the "Vion IMS QTof" menu.
- Select "Reset Instrument" in the "Setup" menu.