LockSpray sample has low sensitivity on Xevo or SYNAPT series MS- WKB19059

Article number: 19059

SYMPTOMS

  • Sensitivity dropoff on LockSpray

ENVIRONMENT

  • Xevo QTof
  • Xevo Tof
  • Xevo G2 QTof
  • Xevo G2 Tof
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si
  • Vion IMS QTof

CAUSE

LockSpray capillary has blown out of the reference probe assembly.

FIX or WORKAROUND

  1. Rebuild LockSpray with ESI reference probe service kit (700004674).

ADDITIONAL INFORMATION