Little or no API gas flow on Quattro LC - WKB17404

Article number: 17404

SYMPTOMS

  • Nebulizer gas flow is lower than expected or shows no flow at all
  • API gas is switched on

ENVIRONMENT

  • Quattro LC
  • Quattro Ultima
  • Quattro II
  • Platform LCZ
  • ZMD
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-MS
  • TQD

CAUSE

The nitrogen tank has run out or the generator is turned off.

FIX or WORKAROUND

Install a new dewar of nitrogen.

ADDITIONAL INFORMATION

Legacy instruments had no nitrogen pressure sensors installed, so there is no alarm when N2 gas pressure starts to drop.