Random data dropout with Xevo G2-S in UNIFI - WKB17462

Article number: 17462

SYMPTOMS

  • During acquisitions, the baseline drops for an interval and there is no data for that time
  • The chromatogram shows a straight line from the point of dropout to the point data resumes
  • Occurs with LC injection and direct infusion with any scan times
  • Error observed in Hyperterminal log: "(wtADCAcqTask): ** Data not transferred"
  • Replacing SATA drive and testing new ADC PCB do not resolve the issue

ENVIRONMENT

  • UNIFI
  • Xevo G2-S

CAUSE

Network cable is plugged into the second port on the instrument LAN card.

FIX or WORKAROUND

Remove the network cable from the instrument LAN card.

ADDITIONAL INFORMATION

dropout_001.png

Not able to find a solution? Click here to request help.