Random data dropout with Xevo G2-S in UNIFI - WKB17462
Article number: 17462
SYMPTOMS
- During acquisitions, the baseline drops for an interval and there is no data for that time
- The chromatogram shows a straight line from the point of dropout to the point data resumes
- Occurs with LC injection and direct infusion with any scan times
- Error observed in Hyperterminal log: "(wtADCAcqTask): ** Data not transferred"
- Replacing SATA drive and testing new ADC PCB do not resolve the issue
ENVIRONMENT
- UNIFI
- Xevo G2-S
CAUSE
Network cable is plugged into the second port on the instrument LAN card.
FIX or WORKAROUND
Remove the network cable from the instrument LAN card.