How much suction or vacuum is needed for the exhaust on a Xevo instrument? - WKB16375

Article number: 16375

ENVIRONMENT

  • Xevo TQD
  • Xevo TQ MS
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo QTof
  • Xevo G2 QTof
  • Xevo G2-S
  • Xevo G2-XS
  • SYNAPT G2
  • SYNAPT G2-S
  • SYNAPT G2-Si

ANSWER

Warning: To avoid exposure to biohazards or toxic LC solvents carried in the nitrogen exhaust, ensure that the nitrogen exhaust is vented via the nitrogen exhaust trap bottle and laboratory exhaust system. The laboratory exhaust system must provide a minimum vacuum of 0.20 kPa (2 mbar, 0.03 psi) below atmospheric pressure (negative pressure).

ADDITIONAL INFORMATION

 

 