Crosstalk due to altered collision cell settings (Xevo G2-XS) - WKB13976
Article number: 13976
SYMPTOMS
- Crosstalk observed in data, crosstalk test fails.
 - LeuEnk spikes into the MSMS sample data when baffle switches, ions are not fully clearing

 - The amount of crosstalk improves with slower scan speeds
 
ENVIRONMENT
- Xevo G2-XS
 - MassLynx
 
CAUSE
Collision cell settings were altered, specifically cell gradient and offset C.

 
FIX or WORKAROUND
- Change collision cell settings back to default values. These settings should not be modified by users. Please check the tech refs for the particular colission cell defaults.
    
- Contact Waters phone support or local field service for assistance.
 
 
ADDITIONAL INFORMATION
id13976, MLYNX, MLYNXV41, SUPMM, XEVOG2XSTF
 