What masses do I enter when infusing QP setup solution, 700003105? - WKB1380
Article number: 1380
ENVIRONMENT
- SQD
- ACQUITY TQD
- Quattro Premier
- Quattro Premier XE
- Quattro micro
ANSWER
For ESI positive, on the tune page, select MS scan in the Function, and then specify the following in the Mass column:
175.1
455.3
1080.8
2034.6
For ESI negative, on the tune page, select MS scan in the Function, and then specify 554 in the Mass column.
ADDITIONAL INFORMATION
The QP solution is available as:
- 700003105 SET-UP SOLUTION QP, SQD
- 700003093 SQD Standards Kit; Vial 1
- 700002646 Waters TQD/Q-Premier/XE/Q-Micro Kit; Vial 1