Source turbo does not reach 100% on QToF; instrument vents - WKB13617
Article number: 13617
SYMPTOMS
- Source turbo reaches only 92%
- Vacuum is fine if the isolation valve is closed
- Backing pressure is high, roughly 4-5 e0
- Instrument vents when isolation valve is opened
ENVIRONMENT
- Xevo G2-XS
- Xevo G2-S
- Xevo G2
- Xevo G3
- SYNAPT G2
- SYNAPT G2-S
- SYNAPT G2-Si
CAUSE
The wrong cone is installed.
FIX or WORKAROUND
- Remove the cone, and pull it out of the sample cone cover.
- Verify that the correct engineering code is imprinted on the cone. Refer to the table below in Additional Information. If not, it is the wrong cone.
- Replace with the proper cone.
ADDITIONAL INFORMATION
Engineering Code |
Part Number |
Instrument |
405006555 |
Xevo QTof |
|
405005883 |
Xevo TQD |
|
405011167 |
SQ Detector 2 |
|
405013693 |
Xevo G2-S QTof |
|
405010077 |
SYNAPT G2-Si |
|
2890101932 |
Xevo TQ-XS |
id13617, XEVOG2SQTF, XEVOG2STOF, XEVOG2XSTF