Error: System store load operation failed - Xevo G2-XS - WKB13595
Article number: 13595
SYMPTOMS
- When selecting File > Load System Settings from Embedded PC (EPC), the following error is seen: System store load operation failed
ENVIRONMENT
- Xevo G2-XS
- Xevo G3
CAUSE
System Settings have not been backed up to the embedded PC (EPC).
FIX or WORKAROUND
Select Backup System Settings (to EPC). The error message will not be seen on subsequent attempts to Load System Settings from EPC.
ADDITIONAL INFORMATION
id13595, XEVOG2XSTF