Skip to main content
Waters

How do I control the MS divert valve in a method in MassLynx? - WKB12652

Article number: 12652

ENVIRONMENT

  • MassLynx
  • SQD
  • 3100
  • SQD2
  • Xevo TQ
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

ANSWER

1.    Go to the Options drop-down menu in the MS Method window and select Method Events.

2.    Create the desired events for flow state to control the divert valve.

3.    Ensure the "Enable" option is selected

clipboard_eaf3a13f1636285f02311baa6fbafc6da.png

ADDITIONAL INFORMATION

 

id12652, A-SQ, A-SQ, MLYNX, MLYNXV41, SQD, SQD2, SUPMM, XEVOTQD, XEVOTQDIVD

Not able to find a solution? Click here to request help.