Error message "Operate trip (software timeout)" - WKB126036
Article number: 126036
SYMPTOMS
- "Operate trip (software time out)"
- Tune page is locked
- Resetting the MS clears the error, but error returns, on average, once a month
ENVIRONMENT
- Xevo TQ S micro
- SQ Detector 2
- MassLynx
CAUSE
This error message typically means the computer has lost communications with the instrument (EPC). The problem could be related to the PC, the cable, or the EPC.
FIX or WORKAROUND
Replace the Ethernet network cable.
ADDITIONAL INFORMATION
This particular investigation was closed after the Ethernet network cable was replaced. The error has not been observed for two months.