Persistent "leak detected error" - WKB125307

Article number: 125307

SYMPTOMS

  • Leak sensor on MS throws "leak detected error"
  • No leak visible in the leak tray
  • Sensor is sprayed with methanol and allowed to air dry
  • Setting changes or reboots do not clear the error

ENVIRONMENT

  • Xevo G2 QTof
  • ACQUITY UPLC I-Class Binary Solvent Manager

CAUSE

Short-circuit in the wiring of the leak sensor

FIX or WORKAROUND

Replace the leak sensor (205000505).

ADDITIONAL INFORMATION

 

id125307, UPAISM, UPBINARY, UPBSM+, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISM, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM, XEVOG2QTOF, XEVOG2TOF