Persistent "leak detected error" - WKB125307
Article number: 125307
SYMPTOMS
- Leak sensor on MS throws "leak detected error"
- No leak visible in the leak tray
- Sensor is sprayed with methanol and allowed to air dry
- Setting changes or reboots do not clear the error
ENVIRONMENT
- Xevo G2 QTof
- ACQUITY UPLC I-Class Binary Solvent Manager
CAUSE
Short-circuit in the wiring of the leak sensor
FIX or WORKAROUND
Replace the leak sensor (205000505).
ADDITIONAL INFORMATION
id125307, UPAISM, UPBINARY, UPBSM+, UPIBSM, UPIBSM, UPIBSM+, UPIBSM+, UPISM, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPPBSM, XEVOG2QTOF, XEVOG2TOF