Skip to main content
Waters

Fluidics system is not infusing sample - WKB125181

Article number: 125181

SYMPTOMS

  • No signal when infusing sample
  • No signal when infusing setup solution
  • Spray is restored when HT diverter valve is switched to LC state (LC must be flowing)

ENVIRONMENT

  • MassLynx
  • Xevo TQ-S
  • Xevo G2-XS QTof
  • Xevo TQ
  • Synapt G2
  • Synapt G2-Si

CAUSE

The bottle reservoir lines are dry or air locked.

FIX or WORKAROUND

  1. Place the fluidics wash line (orange PEEK) in 100% IPA (possibly MS-grade), set fluidics to "Wash", and purge the system several times. Fill a HPLC vial with 100% IPA (possibly MS-grade) and attach it to reservoir A, set fluidics to "Reservoir A", purge the system, and confirm that solvent level drops. Repeat for the remaining reservoir B and C lines if necessary.

ADDITIONAL INFORMATION

If the fix listed above does not help, backflush the reservoir line with the LC flow, connecting with a union the LC tubing and the selected reservoir line tubing.

Alternatively to the LC flow, you can use a syringe (equipped with a blunt needle and a nut and ferrule) for flushing the line and removing the clog or air bubbles.

clipboard_ef524e39c77437af09b8d48eb09bbe59c.png

Not able to find a solution? Click here to request help.