Skip to main content
Waters

API gas shuts off when running detector setup in IntelliStart - WKB12323

Article number: 12323

SYMPTOMS

  • The desolvation gas flow is set too low (see screenshot).
  • API gas is automatically shut off
    clipboard_ef60e8d270aff69691d8cdf6640f8db2e.png

ENVIRONMENT

  • Xevo G2 Tof/QTof
  • Xevo G2-S Tof/QTof
  • Xevo G2-XS Tof/QTof

CAUSE

 

FIX or WORKAROUND

Clear Inlet fatal error and Ext.device error (see screenshots).
image.png
clipboard_e8b3856aea9e72e5a5a358b9a2701b160.png

ADDITIONAL INFORMATION

If there is an error active in the inlet file (e.g. pump over pressure) and the error shutdown is configured, a shutdown is triggered during the detector setup. In these cases it is sufficient to clear the error in the inlet file (LC\Reset Communication) before starting the detector setup.

 

Not able to find a solution? Click here to request help.