Xevo TQ-S micro does not pump down after sample cone is cleaned - WKB122846

Article number: 122846

SYMPTOMS

  • System vents after sample cone is cleaned and turbos do not reach speed when trying to pump down
  • Closing isolation valve before pumping down does not resolve the issue

ENVIRONMENT

  • TQ-S micro
  • Xevo TQD

CAUSE

Failed sample cone and isolation valve O-rings

FIX or WORKAROUND

Replace the sample cone O-ring (700004321) and the isolation valve O-ring (700000902).

ADDITIONAL INFORMATION

 

id122846, XEVOTQSMIC