Skip to main content
Waters

Detector setup fails with beam error - WKB12269

Article number: 12269

SYMPTOMS

  • "Beam error - too many beam intensity failures" or "Beam error - beam too weak" error occurs during detector voltage setup
  • Problem arises even when the beam for LeuEnk is stable
  • Cannot improve beam intensity sufficiently with higher flow rate or sensitivity
  • Abnormal isotopic pattern for Leu Enk (incorrect relative intensity for 2nd and 3rd isotope peaks or artefact peaks are present in between isotopes)
  • 2nd C13 isotopes (M+2) not visible

ENVIRONMENT

  • Xevo G2
  • Xevo G2-S
  • Xevo G2-XS
  • Xevo G3
  • Synapt G2
  • Synapt G2-S
  • Synapt G2-Si
  • Synapt XS

CAUSE

Problem is related to excessively high or low detector voltage at the beginning of the detector setup.

FIX or WORKAROUND

  1. In the Tune page, go to the Setup drop-down menu and select System View
  2. Enter the password: access
  3. Go to the System drop-down menu (not the system 1 or system 2 tabs)
  4. Select Auto Values and adjust the detector voltage in Resolution Mode manually
    clipboard_e152f3af14d221821be2cb3efb6c3c160.png
  5. If the signal is too high and shows artefact peaks after the real ones, decrease the detector voltage to remove artefacts but ensure LeuEnk is still giving a strong signal.
  6. If the signal is too low and the isotopic pattern of Leu Enk is abnormal, increase the detector voltage to reach signal in mid e5
    clipboard_ebe7a700821d3c80a5f96769f47d7ca04.png 
  7. Click Update and Close.
  8. Save Tune Page with system Settings an backup to EPC
    clipboard_ed61e25ea3d198927b7188e7aa8906877.png
  9. Run Detector Setup.
  10. Repeat for Negative Ion mode, if necessary and backup the system settings to EPC once more to save the optimal values

ADDITIONAL INFORMATION

Make sure that the beam is stable (too high desolvation temp. and gas flows can cause an unstable beam)

Nominal mass position should be adjusted as needed

Capillary voltage should not be too high as this may cause overloading of the detector

Rapid changes in detector gain are typically seen with new detectors or when the instrument has recently been switched back on after several months of inactivity

id12269, MALDIG2SI, SYNAPTXS, SYNG2HD, SYNG2HDMAL, SYNG2MS, SYNG2MSMAL, SYNG2SHD, SYNG2SHDML, SYNG2SIHD, SYNG2SIMS, SYNG2SMS, SYNG2SMSML, XEVOG2QTOF, XEVOG2SQTF, XEVOG2STOF, XEVOG2TOF, XEVOG2XSTF, XEVOG3QTOF

Not able to find a solution? Click here to request help.