The PEEK tubing from the ACQUITY UPLC system leaks at the LC port on the diverter valve for any Xevo Mass Spec - WKB122296
Article number: 122296
SYMPTOMS
- The PEEK line fitted to the LC port on the diverter valve pops out of the fitting when flow is on
- Replacing the probe assembly for the Tool-Free ESI probe does not resolve the issue
ENVIRONMENT
- Xevo TQ-S micro
- Xevo TQ-S
- Xevo TQ MS
- Xevo TQD
- Xevo G2-S
- Xevo G2-XS
- Xevo TQ-XS
- Tool-Free ESI probe
CAUSE
Blockage between the LC port and the Source port on the diverter valve
FIX or WORKAROUND
- Disconnect the probe assembly from the Source port on the diverter valve.
- Disconnect the PEEK tubing from the LC port on the diverter valve.
- Connect the PEEK tubing from the LC port to the Source port on the diverter valve.
- Ensure that the flow state for the diverter valve is set to LC on the Fluidics tab of the tune page.
- Start the flow to back flush the ports.
ADDITIONAL INFORMATION