Low sensitivity with Xevo TQD when running injections or infusing sample - WKB121126
Article number: 121126
SYMPTOMS
- Low sensitivity when running injections or infusing sample
ENVIRONMENT
- Xevo TQD
CAUSE
The capillary voltage readback does not match the set capillary voltage.
Putting the system in Standby drops the capillary readback to 0 kV.
The electronics are not controlling the capillary PSU normally; other user-adjustable voltages can be controlled normally.
FIX or WORKAROUND
- Close MassLynx and reboot the EPC using either telnet or the reset switch, which is just to the right of the power switch on the front panel.
- After the EPC reboots (wait approximately 2 minutes), start MassLynx and verify that the capillary voltage readback is now correct and matches the set capillary voltage.
ADDITIONAL INFORMATION
id121126, XEVOTQD, XEVOTQDIVD