Low sensitivity on Quattro Premier - solution 2 - WKB121134

Article number: 121134

SYMPTOMS

  • Low sensitivity on Quattro Premier
  • Drop in sensitivity depends on analyte

ENVIRONMENT

  • Quattro Premier
  • Quattro Premier XE

CAUSE

The ion source enclosure is not saturated with mobile phase adducts, a condition that impedes ionization.

FIX or WORKAROUND

  1. The source enclosure requires passivation with mobile phase additives.
  2. Set the high flow from UPLC with initial gradient composition for a minimum of 30 minutes to passivate the source enclosure. API and collision gas must be ON.

ADDITIONAL INFORMATION

 

id121134, eluent