Xevo G2-XS fails calibration - WKB119723
Article number: 119723
SYMPTOMS
- Failure of calibration protocol
- Beam attenuation is less than 1% during calibration setup
- "!" seen on calibration data report
ENVIRONMENT
- Xevo G2-XS
CAUSE
A high-intensity contaminant peak in the calibration solution causes calibration peaks to be severely attenuated.
FIX or WORKAROUND
- Resolve the issue with calibration solution (remake the solution or install a new solution on the fluidics reservoir).
- Calibrate the system and ensure that the calibration passes.
ADDITIONAL INFORMATION
id119723, XEVOG2XSTF