Low sensitivity after a PM - WKB116099
Article number: 116099
SYMPTOMS
- Low sensitivity after a PM
- Setup solution looks typical
- Only some of the customer's analytes are affected
- Correctly linked calibration and resolution settings
ENVIRONMENT
- MassLynx
- Quattro Premier
- Quattro Premier XE
- Quattro LC
- Quattro LCZ
- Quattro micro API
- Quattro Ultima
- 3100 Mass Detector
- RADIAN ASAP Direct Mass Detector
- ACQUITY SQD system
- SQ Detector
- ACQUITY TQD system
- TQ Detector
- SQ Detector 2
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-GC
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
- ZQ 2000
- ZQ 4000
- ZMD
CAUSE
During the PM, the customer's method was written over.
FIX or WORKAROUND
Open pre- and post-PM experimental records. Ensure that the parameters for the customer's method (desolvation temperature, gas flow, and so on) match.
ADDITIONAL INFORMATION
id116099, MLYNX, MLYNXV41, SUPMM, XEVOTQD, XEVOTQDIVD